Guidance

Field safety notices: guidance for manufacturers

How to write clear and effective field safety notices (FSNs) for medical devices.

Manufacturers must take field safety corrective actions (FSCA) to prevent or reduce the risk of a serious incident in relation to a device which has already been placed on the market or put into service.

You should communicate details of FSCA to customers and users using a field safety notice (FSN). The UK medical devices regulations state that after 바카라 사이트submitting the initial report and proposed notice바카라 사이트 the manufacturer must implement the FSCA as soon as possible and monitor its progress.바카라 사이트

You can find out more about writing and distributing FSNs in our guidance on post-market surveillance.

This guidance for manufacturers provides supplementary information and advice on how to produce and distribute effective FSNs.

Good traceability

You should keep records to help trace your distributed product. This should include:

  • records of medical devices by manufacturing date and batch, version or serial number

  • unique device identifiers (UDI)

  • keeping traceability of medical devices directly supplied to users and distributors

Contracts with distributors should include keeping onward traceability records to end users and agreements on information sharing as far as is practicable. If distributors are not willing to share customer lists with you, your contract should require them to do so directly with the MHRA on request. Responsibility for ensuring the communication reaches end users remains with the manufacturer.

For connected devices, software and applications (apps), consider:

  • how you can use the technology to aid traceability

  • how to distribute the message of FSNs

  • how to enable updates to be delivered to the field

When supplying to large organisations such as NHS trusts, recording the address and name of the responsible department is important for traceability. Where possible, record the name of the user or person responsible for the device as well as anybody responsible for the purchase.

In addition, when supplying NHS organisations, recording the name of the NHS trust will enable you to identify the medical device safety officer (MDSO) to contact. This will help you target your communication to get maximum replies to your FSNs.

FSN바카라 사이트cDzԳٱԳ

Manufacturers can use this .Where a section of the template does not apply to your product, you can remove it.

You should read the바카라 사이트 before completing your notice.

There is also an바카라 사이트 Ի바카라 사이트a .

The FSN should be on a company letterhead, be written in English and include:

  • a clear title, with 바카라 사이트Urgent FIELD SAFETY NOTICE바카라 사이트 followed by the commercial name of the affected product, an FSCA identifier (for example, date or FSCA reference number) and the type of action
  • specific details to enable the affected product to be easily identified, for example, type of device, model name and number, batch/lot or serial numbers of affected devices, part or order number and UDIs where available - we recommend that the manufacture dates are also included for ease of identification
  • a factual statement explaining the reasons for the FSCA, including a description of the device deficiency or malfunction, clarification of the potential hazard associated with the continued use of the device and the associated risk to the patient, user or other person and any possible risks to patients associated with previous use of affected devices.
  • advice on actions to be taken by the user, including:
    • how to identify an affected device
    • method of recovery, quarantine, disposal or modification of device
    • review and implementation of additional risk mitigation measures
    • recommended review of patient바카라 사이트s previous results or patient follow up, for example, implants, IVD
    • timelines for completion of these actions
  • a request to pass the field safety notice to all those who need to be aware of it in the organisation and to maintain awareness over an appropriate defined period
  • if relevant, a request for a copy of the field safety notice to be passed on to all organisations to which any affected devices have been transferred, and for the details of onward distribution of these devices to be given to the manufacturer
  • if relevant, a request that the recipient of the field safety notice alerts other organisations to which incorrect test results from the use of the devices have been sent, such as failure of diagnostic tests
  • confirmation that the MHRA has been advised of the FSCA
  • contact point for customers how and when to reach the designated person. This person should be either the manufacturer if UK based, the UK responsible person or Northern Ireland authorised representative
  • an acknowledgment form for the receiver

Do not include any comments and descriptions that attempt to:

  • play down the level of risk in an inappropriate manner
  • advertise products or services

Maximising efficacy of FSN dzܲԾپDz

Write the바카라 사이트FSN바카라 사이트in a clear style and at an appropriate level for the intended audience. Remember that in some cases patients will read the notice, so avoid jargon or unnecessary technical language.

The risk to the user is the key element of the notice. Make it clear what the problem is and what the recipient must do.

Do not delay sending an바카라 사이트FSN바카라 사이트because you are waiting to include information on the cause of the problem. In such circumstances, there is the option to send a follow-up바카라 사이트FSN.

The FSN must be sent in a format that enables customers to search for UDI information (wherever available), catalogue numbers, model numbers and LOT numbers. We consider it best practice to include unique device identifiers (UDIs) in your field safety corrective action (FSCA) communications.

We have worked with바카라 사이트바카라 사이트aԻ바카라 사이트바카라 사이트to produce the following templates for the incorporation of detailed device information in field safety notices (FSNs) and we encourage you to use these.

If you use GS1 UDIs, use the바카라 사이트바카라 사이트. ճ바카라 사이트바카라 사이트provides background information and details on how to fill in the spreadsheet.

If you use HIBCC UDIs, use the 바카라 사이트바카라 사이트.

Manufacturers must submit their initial FSCA report and draft바카라 사이트FSN바카라 사이트to us before sharing with customers (except for urgent FSNs). This is so we can provide advice on the FSCA implementation strategy or comments on the proposed FSN. If you receive no response within 5 days of submitting the draft, you should circulate the FSN to your customers. If the action required to protect patient safety is urgent, you can circulate the FSN to your customers without waiting the 5 days for a response from MHRA.

You should issue the바카라 사이트FSN, with the crucial acknowledgement instructions, as soon as possible after the final version has been agreed.

These are some of the common problems in FSNs:

  • missing lot numbers

  • too much jargon

  • badly explained or missing instructions on what to do next

  • missing detailed description of the problem

  • missing explanation of how the problem affects the patient

  • incorrect or missing contact details for the UK

  • missing acknowledgment form

  • ھԲ바카라 사이트FSN바카라 사이트has not been signed or dated

  • poor translations 바카라 사이트 we suggest asking your UK responsible person or Northern Ireland authorised representative, or importer or distributor in the UK to check the Englishif this could be an issue

In your final version바카라 사이트FSN we are looking for the following:

  • the correct contact details for the UK

  • only generic information about the recipients - do not use a sample with an individual바카라 사이트s name or address on it

  • include covering letters in the바카라 사이트FSN

  • send it to us as a single PDF file with all the relevant information rather than separate files with versions of FSNs (such as one aimed at distributors and another at customers)

  • the PDF must not be a scanned version of a paper document because it cannot be searched for serial numbers, model numbers and similar information

  • the UDIs of the affected devices, preferably on the바카라 사이트UDI바카라 사이트device and header Excel spreadsheet

  • if possible, include NHS supply chain codes to help your customers decide if they have your product in stock

  • if the바카라 사이트FSCA바카라 사이트covers all lots, state this as well as listing all the numbers

We publish weekly lists of FSNs that affect the UK which are seen by over 10,000 subscribers.

Effective targeting of FSNs and maximising response rates

The UK medical devices regulations state that 바카라 사이트the manufacturer must take all reasonable steps to ensure the field safety notice is sent to users of the device바카라 사이트.

The MHRA does not stipulate the method of distributing FSNs. The key element in maximising response rates is to make sure the right people know that the바카라 사이트FSN바카라 사이트is urgent and needs a response.

A 바카라 사이트read receipt바카라 사이트 of an email is not acceptable proof of acknowledgement of an바카라 사이트FSN바카라 사이트because it does not prove that the바카라 사이트FSN바카라 사이트has been read and acted on. Some companies are now using an e-signature portal as confirmation that FSNs have been received, and acted on, by the organisation. We encourage this route of distributing FSNs.

Registered post only proves that the바카라 사이트FSN바카라 사이트arrived at its destination, not that it has been read and acted on by the intended recipient.

Using the network of medical device safety officers (MDSO) in healthcare establishments throughout England can help you get replies to your FSNs. You can access the list of contacts is through an account with our바카라 사이트. With a MORE account, you can submit and manage your reports online, or just use the바카라 사이트MDSO바카라 사이트email lists to send FSNs. The list is updated monthly and includes the contact details for Scottish equipment co-ordinators.

MDSOs must be registered with the MHRA and part of their role is to act as an additional senior point of contact for manufacturers and support local actions on FSNs. Although not all healthcare organisations have an바카라 사이트MDSO, over 95% of NHS trusts do, so it is a great advantage to include MDSOs when sending FSNs to your customers.

The MHRA has a바카라 사이트flyer that explains FSNs바카라 사이트 and why they are important, which manufacturers can send to their customers to help achieve target reconciliation on responses.

The manufacturer바카라 사이트s바카라 사이트FSCA바카라 사이트strategy

ճ바카라 사이트FSCA바카라 사이트strategy will need to specify whether a general public warning (advertisement, a helpline number for 24 hour access to further information or media release) is needed and whether to use more specialised news media. The latter can allow targeting of specific segments of the population (for example, customers signed up to receive updates from manufacturer websites or loyalty schemes) to prevent unnecessary public anxiety and advise that consultation between patients and their healthcare professional is essential.

Whichever method is used, manufacturers should inform the MHRA first and send us a draft copy for comment. There is advice on advertising and using the media in the PROSAFE바카라 사이트.

For FSNs relating to software devices published on app stores, a summary of the FSN should be displayed in the information on the app store and, where possible, at the app바카라 사이트s start-up. If the full FSN cannot be displayed, a link to the full text should be made available in-app. It is not acceptable to fix a software issue without issuing an FSN to explain the action taken.

Confidentiality of FSNs

The MHRA keeps information from manufacturers on proposed field safety corrective actions (FSCAs), including draft copies of바카라 사이트FSN, confidential as required by the Medical Devices Regulations (SI 2002 No 618, as amended) (UK MDR 2002). Once a manufacturer has issued a ھԲ바카라 사이트FSN바카라 사이트to customers, it is no longer treated as confidential because it is in the public domain.

The MHRA has agreements in place with other international regulatory authorities, where we circulate notifications about FSCAs (in confidence) to promote co-ordinated actions between international regulatory authorities.

What the MHRA does when there is an바카라 사이트FSCA

Where there is doubt, the MHRA will provide its interpretation of whether the action proposed by the manufacturer falls within the definition of an바카라 사이트FSCA바카라 사이트and will give advice on FSNs and associated바카라 사이트FSCA바카라 사이트strategies.

As soon as possible after we receive sufficient information from the manufacturer, we review the details of every바카라 사이트FSCA바카라 사이트. We expect manufacturers to reply quickly to queries about FSCAs. Delayed replies increase the likelihood of the MHRA publishing separate advice to the health service in the interests of protecting public health. It may also lead to the MHRA initiating legal compliance action.

Circumstances where the MHRA might have to issue separate advice include the need to:

  • supplement information provided by the manufacturer, for example, when the message of the바카라 사이트FSN바카라 사이트is not clear or where additional advice is needed

  • bring the바카라 사이트FSCA바카라 사이트to the attention of a wider user base than that contacted directly by the manufacturer, for example, to target different or additional professionals

  • notify chief executives of NHS trusts or other management personnel, for reasons of clinical governance, of information being issued directly to healthcare professionals

  • issue an MHRA statement where a safety issue has sufficiently high profile that the health service would expect it

  • help ensure the message gets to everyone affected when a very large number of customers and centres are affected, where there is a possibility that devices may have been moved between healthcare providers or centres without the knowledge of the manufacturer or distributor

  • give different advice to that provided by the manufacturer, although through negotiation we try to avoid this wherever possible

The manufacturer, or their authorised representative, will almost always get the opportunity to comment on the draft of any safety-related notice that the MHRA produces.

For FSCAs with public, media or a particular health interest, the MHRA will liaise with colleagues in the Department of Health and Social Care and with the manufacturer to prepare information for general release. Media statements should be worded to minimise any public alarm. Where necessary, UK health ministers will be kept informed.

The MHRA will agree appropriate milestones with the manufacturer or authorised representative for receiving바카라 사이트FSCA바카라 사이트status reports, including a final report.We examine all the reports from the manufacturer and keep an ongoing assessment of the effectiveness of the바카라 사이트FSCA바카라 사이트action. This may include further referral to the weekly management meeting.

Where an바카라 사이트FSCA바카라 사이트is initiated following a report of an incident submitted to the MHRA, we will communicate the outcome of the investigation to the person who sent us the report. In the interests of worldwide patient and user safety, we may also provide copies of MHRA safety warnings issued to the health service in the UK to other regulatory authorities.

Our FSN바카라 사이트review and monitoring process

  1. FSCA and FSN바카라 사이트logged and initial email sent to manufacturer from AIC requesting information about the action being taken.

  2. FSCA assigned to benefit risk evaluation assessor (BRE assessor), who conducts initial review of safety message and required actions. Therapeutic area group meeting reviews바카라 사이트FSCA where required.

  3. BRE assessor may ask for more information from manufacturer or seek clinical opinion.

  4. Information on reconciliation and long-term corrective and preventive action (CAPA) received from manufacturer in response to initial email from AIC, and/or request from BRE assessor.

  5. BRE assessor completes assessment including recommendations for MHRA actions relating to the바카라 사이트FSCA.

  6. BRE assessor completes the actions agreed, for example, continue to monitor, or publish safety communications.

Field corrective actions

Although they are outside the scope of the notification requirements of the UK MDR 2002, the MHRA encourages manufacturers to deal with field corrective actions (FCAs) in a similar way to FSCAs.

A field corrective action may arise from a more minor issue that does not pose a risk of death or serious injury.

Examples of field actions considered to be FSCAs

Example 1

A manufacturer identifies that water used in the manufacturing process is contaminated, and all devices manufactured using this process are potentially contaminated. The investigation included consideration of the type of contamination, the causes of the contamination, and the impact use of a contaminated device could have, taking into account different users.

Rationale: there is potential for serious injury should a contaminated device be used.

Example 2

A manufacturer identifies that diagnostic kit for detecting antibodies impacting on fertility processes within human semen or serum has the potential to produce false negative results.

Rationale: any corrective action(s) taken by the manufacturer to reduce the risk of direct or indirect harm associated with the use of a 바카라 사이트defective바카라 사이트 diagnostic kit should be reportable. This is especially true in the cases where patient follow up measures are required.

Example 3

The manufacturer updates the instructions for use to include a new warning to mitigate a serious risk to the user.

Rationale: the purpose of the new warning is to prevent serious injury to a patient or user.

Example 4

Following reports that a Heaf test seemed to give an abnormally high level of 바카라 사이트false negative results바카라 사이트, the manufacturer identified that, although it was performing to specification and requirements, the firing mechanism could be interrupted if it was not handled in a specific way. The instructions for use supplied with the device did not clearly identify this handling requirement. The manufacturer changed the instructions so that the handling of the unit was clearly highlighted in both text and diagram form.

Rationale: there is potential for misfired units to give rise to false negatives, which in turn could result in inappropriate tuberculosis immunisation.

Example 5

Sterilisation services are provided for a manufacturer by a third-party supplier. Devices are found not to have been sterilised in line with the documented, validated process. The investigation included consideration of the impact use of an unsterile device may have as well as the impact removal of product from the market would have on supply.

Rationale: there is potential for serious injury should an affected device be used.

Example 6

A manufacturer of a syringe pump identified a small risk that pumps within a range of serial numbers may not alarm if the syringe plunger clamp was left open. The manufacturer issued instructions on detecting and correcting the problem. Instructions on checking the 바카라 사이트Clamp Open바카라 사이트 detection mechanism during routine maintenance were also added to the service manual.

Rationale: there is potential for serious injury as a result of over- or under-infusion.

Example 7

A manufacturer identified that due to a battery defect, the rate of battery depletion towards the end of service life of one of their pacemaker models was more rapid than originally anticipated through accelerated testing. There were no un-implanted units remaining with distributors or in hospital supplies available for return to the manufacturer. The manufacturer issued written advice to clinicians following patients implanted with these pacemakers, emphasising the need to schedule clinic visits more frequently than indicated in the physicians바카라 사이트 manual supplied with the product, to check the pacemaker battery status.

Rationale: failure to detect early signs of battery depletion would risk the patient losing pacing therapy.

Example 8

A test for detecting bacterial antigen in cerebrospinal fluid is found to cross-react with another bacterium which causes meningitis.

Rationale: this could result in the wrong antibiotics being administered.

Example 9

Following reports of users falling from powered wheelchairs due to failures of castor assemblies it was found that the instructions for use did not include adequate user checks and regular maintenance requirements to ensure that the castors could continue to operate correctly. The manufacturer revised the instructions for use and incorporated new requirements for user functional checks and regular maintenance.

Rationale: there is potential for serious injury as the result of a fall.

Examples of field actions not considered to be FSCAs

Example 1: general medical device

A manufacturer places an incorrect expiry date of 18 months on the labelling of a batch of product. The supported shelf life is 2 years. The manufacturer chooses to send new labels with the correct expiry date to customers.

Example 2: out-of-box failure

A product is shipped with a reagent missing. Users cannot run the test without it. The manufacturer exchanges distributed product for complete test kits.

Updates to this page

Published 20 May 2020
Last updated 15 January 2025 show all updates
  1. Updated to reflect the laying of The Medical Devices (Post-market Surveillance requirements) (Amendment) Regulations 2024.

  2. This has been updated due to the end of the Brexit transition period.

  3. First published.

Sign up for emails or print this page