Oxylog 3000 and Oxylog 3000 plus ventilator 바카라 사이트“ risk of failure
Manufactured by Dräger 바카라 사이트“ ventilator may stop working if oxide layers accumulate due to the infrequent use of the control knobs (potentiometers).
Action
-
Identify all Oxylog 3000 and Oxylog 3000 plus ventilators.
-
Ensure all users are aware of the and know how to perform the pre-use check outlined in the instructions for use (IFU).
-
Contact Dräger to acknowledge the FSN and schedule a time to perform the software update.
-
Update the IFU with the supplement provided with the FSN or by Dräger바카라 사이트™s service engineers at the time of software installation.
Action by
All those who use these devices.
Deadlines for actions
Actions underway: 1 March 2017
Actions complete: 15 March 2017
Problem / background
This error condition (referred to by Dräger as 바카라 사이트˜Poti unplugged바카라 사이트™) which could result in the ventilator stopping, is caused by an accumulation of an oxide layer in the potentiometer when these are rarely (or never) moved. The layer develops over time, but twisting the potentiometer knobs prevents this, as described in that was sent to customers in 2015.
Dräger have now issued software which enables the ventilation to continue with the last, valid parameters (and posts the corresponding alarms). Additionally, the new software prompts turning of the potentiometer knobs in the pre-use device check, so the issue should not occur.
Manufacturer contacts
Dräger Quality Department
Tel: 01442 292 870
Email Helen.Glass@draeger.com
Distribution
If you are responsible for cascading these alerts in your organisation, these are our suggested distribution lists.
Trusts (NHS boards in Scotland)
CAS and SABS (NI) liaison officers for onward distribution to all relevant staff including:
- A&E departments
- Adult intensive care units
- Ambulance staff
- Anaesthetic medical staff
- Biomedical engineering staff
- EBME departments
- Equipment stores
- Equipment libraries and stores
- Health and safety managers
- Intensive care units
- Medical physics departments
- Paediatric intensive care units
- Paediatric wards
- Patient transport managers
- Purchasing managers
- Resuscitation officers and trainers
- Risk managers
- Special care baby units
Independent distribution
Establishments registered with the Care Quality Commission (CQC) (England only):
- Hospitals in the independent sector
Please note: CQC and OFSTED do not distribute these alerts. Independent healthcare providers and social care providers can sign up to receive MDAs directly from the Department of Health바카라 사이트™s Central Alerting System (CAS) by sending an email to: safetyalerts@dh.gsi.gov.uk and requesting this facility.
Enquiries
England
Send enquiries about this notice to MHRA, quoting reference number MDA/2017/001 or 2016/012/012/291/014.
Technical aspects
Hiten Patel or Emma Rooke, MHRA
Tel: 020 3080 6115 / 6609
Email: hiten.patel@mhra.gsi.gov.uk or emma.rooke@mhra.gsi.gov.uk.
Clinical aspects
MHRA Devices Clinical Team
Tel: 020 3080 6000 (Customer Services)
Email: DCT@mhra.gsi.gov.uk
Reporting adverse incidents in England
Northern Ireland
Alerts in Northern Ireland are distributed via the
Enquiries and adverse incident reports in Northern Ireland should be addressed to:
Northern Ireland Adverse Incident Centre,
CMO Group,
Department of Health,
Social Services and Public Safety
Tel: 028 9052 3868
Email: niaic@health-ni.gov.uk
Web:
Reporting adverse incidents in Northern Ireland
Please report directly to NIAIC using the forms on our
Scotland
Enquiries and adverse incident reports in Scotland should be addressed to:
Incident Reporting and Investigation Centre,
Health Facilities Scotland,
NHS National Services Scotland
Tel: 0131 275 7575
Fax: 0131 314 0722
Email: nss.iric@nhs.net
Reporting adverse incidents in Scotland
NHS Boards and Local Authorities in Scotland 바카라 사이트“ report to
Contractors such as private hospitals carrying out NHS work and private care homes that accept social work funded clients 바카라 사이트“ report to
Private facilities providing care to private clients report to the and
Wales
Enquiries in Wales should be addressed to:
Healthcare Quality Division,
Welsh 바카라 사이트
Tel: 02920 823 624 / 02920 825 510
Email:Haz-Aic@wales.gsi.gov.uk
Reporting adverse incidents in Wales
Report to MHRA through and follow specific advice for reporting in Wales in
Download document