Alaris® syringe pumps (GH, CC, TIVA & PK models) 바카라 사이트“ risk of uncontrolled bolus of medicine.

[Archived] Manufactured by CareFusion 바카라 사이트“ Updates and replaces MDA/2016/023 as the manufacturer has now identified the potential for unintended bolus of medication when any syringe type is used in a pump that contains a broken spring in the plunger assembly.

Update

This MDA was superseded by MDA/2017/006 in April 2017.

Archive

Updates to this page

Published 20 February 2017